6
OPERATION
98
IFC 300
www.krohne.com
06/2009 - 4000069802 - MA IFC 300 R03 en
No.
Function
Settings / descriptions
C1.3 self test
C1.3
self test
C1.3.1
empty pipe detect
Switch conductivity measurement off and on (measurement of the electrode
resistance).
Select: off
conductivity (only conductivity measurement)
cond. + empty pipe [F] (conductivity measurement and empty pipe indication,
error category [F] application);
Flow indication "= 0" when pipe empty
cond. + empty pipe [S] (conductivity measurement and empty pipe indication,
error category [S] measurement outside of specification);
Flow indication "
≠
0" when pipe empty
C1.3.2
limit empty pipe
Only available when empty pipe activated [..] in Fct. C1.3.1
Range: 0.0
…
9999
μ
S (set max 50% of the lowest occurring conductivity in
operation. Conductivity below this value = signal as empty pipe)
C1.3.3
act. conductivity
Only available when empty pipe activated [..] in Fct. C1.3.1
Current conductivity is indicated. Activation takes place only after setting
mode is exited!
C1.3.4
full pipe detect
Only for measuring sensors with 3 (4) electrodes.
Select: off (no full pipe measurement) / on (full pipe measurement by 3rd
electrode)
C1.3.5
limit full pipe
Only when full pipe detection activated, see Fct. C1.3.4.
Range: 0.0
…
9999
μ
S (conductivity over this value = signal as full pipe)
C1.3.6
linearity
Only if GK values "GK+GKL" are activated with Fct. C1.1.3 (check carried out
with 2 field currents).
Select: off (no linearity check) / on (linearity check activated)
C1.3.7
act linearity
Only available when linearity test "on" activated in Fct. C1.3.6. The
conductivity measurement must also be activated, see Fct. C1.3.1.
Activation takes place only after setting mode is exited!
C1.3.8
gain
Automatic test switched off / on, select: off / on
C1.3.9
coil current
Automatic test switched off / on, select: off / on
C1.3.10
flow profile
Automatic test switched off / on, select: off / on
C1.3.11
limit flow profile
Only with flow profile switched on, see Fct. C1.3.10.
Range: 0.000
…
10 (absolute values above this threshold generate an error of
category [S])
C1.3.12
act. flow profile
Only available when flow profile "on" activated in Fct. C1.3.10. Activation
takes place only after setting mode is exited!
C1.3.13
electrode noise
Automatic test switched off / on, select: off / on
C1.3.14
limit electr. noise
Only with electrode noise activated, see Fct. C1.3.13.
Range: 0.000
…
12 m/s (noise above this threshold generates an error of
category [S])
C1.3.15
act. electr. noise
Only available when electrode noise "on" activated in Fct. C1.3.13. Activation
takes place only after setting mode is exited!
C1.3.16
settling of field
Automatic test switched off / on, select: off / on
.book Page 98 Wednesday, June 10, 2009 4:20 PM