ELECTRICAL CONNECTIONS
4
77
IFC 300
www.krohne.com
06/2009 - 4000069802 - MA IFC 300 R03 en
4.11.5 Exi inputs/outputs
Current output active (only current output terminals C/C- have HART
®
capability), Exi
I/Os
•
Observe connection polarity.
•
Uint,nom = 20 VDC
•
I
≤
22 mA
•
R
L
≤
450
Ω
•
X designates the connection terminals A or C, depending on the version of the signal
converter.
Current output passive (only current output terminals C/C- have HART
®
capability), Exi
I/Os
•
Any connection polarity.
•
U
ext
≤
32 VDC
•
I
≤
22 mA
•
U
0
≥
4 V
•
R
L, min
= (U
ext
- U
0
) / I
max
•
X designates the connection terminals A or C, depending on the version of the signal
converter.
DANGER!
For devices used in hazardous areas, additional safety notes apply; please refer to the Ex
documentation.
INFORMATION!
For further information on electrical connection refer to Description of the inputs and outputs on
page 55
.
Figure 4-43: Current output active I
a
Exi
Figure 4-44: Current output passive I
p
Exi
.book Page 77 Wednesday, June 10, 2009 4:20 PM