OPERATION
6
55
H250 M40
www.krohne.com
03/2020 - 4000640706 - MA H250-M40 R06 en
1.4 DISPLAY
1.4 DISPLAY
1.4 DISPLAY
1.4 DISPLAY
Different measured values can be selected for permanent or alternating display.
The reading on the display can be rotated.
1.4.1
MEAS
’
D VALUE
Permanent display of flow in flow units
COUNTER
Permanent display of totalizer
MV/COUNTER
Alternating display of flow value in flow units and totalizer
MV&COUNTER
Simultaneous display of flow value and totalizer
PERCENT
Permanent display of flow value in percent
1.4.2 ROTATION
0
°
The reading is not rotated.
90
°
The reading on the display is rotated by 90
°
.
180
°
The reading on the display is rotated by 180
°
.
270
°
The reading on the display is rotated by 270
°
.
1.5 TIME CONST
1.5 TIME CONST
1.5 TIME CONST
1.5 TIME CONST
0.0...20.0s
Given in seconds
The output variables (value of the current loop and indicated flow value) follow the current
process by the value set here (in seconds) with a time delay.
Note:
Note:
Note:
Note:
If the current flow is polled via HART
®
communication, the transferred measured value is
also played back with a delay.
1.6 RESET
1.6 RESET
1.6 RESET
1.6 RESET
Local resetting of totalizer and acknowledgement of warnings.
There is always a safety prompt ("YES / NO"), to avoid accidental resetting.
Note:
Note:
Note:
Note:
External resetting of the totalizer can be installed with the binary input B3.
1.6.1 COUNTER
Confirming with "YES" resets the value of the totalizer to 0.0.
1.6.2 ERROR
Confirming with "YES" acknowledges all existing warnings.
Note:
Note:
Note:
Note:
Confirming with "YES" acknowledges existing errors and warnings.
1.6.3 RE-INIT IO
As a general rule, data is transferred from the counter module to the ESK4 and vice versa
when the device starts. For safety, another data transmission can be performed by
selecting this menu item and confirming with "YES".
Table 6-12: Menu description - 1 OPERATION