03/2020 - 4000640706 - MA H250-M40 R06 en
1.2 OUTPUT B2
1.2 OUTPUT B2
1.2 OUTPUT B2
1.2 OUTPUT B2
Output B2 is a binary switching output. In Fct. 3.4, one of the following functions can be
selected for this output:
INACTIVE, SWITCH POINT, COUNT. LIMIT, PULSE OUTP.
One of the following functions can be selected as contact type in Fct. 3.5:
NORM.OPEN
1
/ NORM.CLOSED
2
1.2.1
INACTIVE
-
MEAS.VAL. B2
Refer to MEAS.VAL. B1
In Fct. 1.2.2 a hysteresis can be specified additionally.
CNT. VAL. B2
Refer to CNT. VAL. B1
PULSE WIDTH B2
Refer to PULSE WIDTH B1
A configuration takes place in Fct. 1.3.1 PULSE WIDTH, 1.3.2 PULSE/UNIT and 3.11.2
COUNTER.
1.2.2 HYST. B2
Refer to HYST. B1
1.3 PULSE OUTP.
1.3 PULSE OUTP.
1.3 PULSE OUTP.
1.3 PULSE OUTP.
-
1.3.1 PULSE WIDTH
50ms
T
i
= 50 ms; f
max
= 10 Hz
max. pulses/h = 36000
100ms
T
i
= 100 ms; f
max
= 5 Hz
max. pulses/h = 18000
200ms
T
i
= 200 ms; f
max
= 2.5 Hz
max. pulses/h = 9000
500ms
T
i
= 500 ms; f
max
= 1 Hz
max. pulses/h = 3600
1.3.2 PULSE/UNIT
0.001...1000
Pulse per volume or mass unit of counter (setting in Fct. 3.11.2), which can be output via
one of the binary outputs. The maximum frequency for the pulse output (refer to Fct 1.3.1)
cannot be exceeded, even at maximum flow rate (full scale).
Example:
Example:
Example:
Example:
Final value Q
max
= 1200 L/h; volume unit of counter = Litre; pulse width = 100ms;
If the factor 1 is entered, 1 pulse/litre = 1200 pulses are generated in one hour at maximum
flow rate.
Number of max. permissible pulses:
Selection / Input
Explanation