17
Pipe wall thickness now appears on the display.
After entering the pipe wall thickness, press ENTER.
QUICK START yy-mm-dd hh:mm:ss
Dimension units MILLIMETRES
Pipe outside diameter? 58.0
Pipe wall thickness? 4.0
Pipe lining thickness now appears on the display. If the pipe you
are measuring has a lining, you now enter the Pipe lining
thickness. If nothing is entered the instrument automatically
assumes there is no lining, press ENTER to move on. If the
application has a pipe lining, enter the required thickness in the
units selected. Press ENTER to continue.
QUICK START yy-mm-dd hh:mm:ss
Dimension units MILLIMETRES
Pipe outside diameter? 58.0
Pipe wall thickness? 4.0
Pipe lining thickness?
0.0
The instrument now displays Select pipe wall material. By using
the scroll keys it is possible to scroll up or down the options
available. Select the required material and press ENTER.
QUICK START yy-mm-dd hh:mm:ss
Select pipe wall material:
Mild Steel
S’ less Steel 316
S’ less Steel 303
Plastic
Cast Iron
Ductile Iron
Copper
Brass
Concrete
Glass
Other (m/s)
The following will only be displayed at this stage if a lining
thickness had been entered previously.
Use the scroll keys to select the required material then press
ENTER. If Other is selected, enter the propagation rate of the
lining in metres/sec. Contact KROHNE if this is not known.
QUICK START yy-mm-dd hh:mm:ss
Select pipe lining material:
Steel
Rubber
Glass
Epoxy
Concrete
Other (m/s)
Select fluid type now appears on the display.
Use the scroll keys to select the application liquid and press
ENTER. If Measure is selected, the instrument automatically
measures the propagation rate of the liquid, but only when the pipe
size is greater then 40 mm internal diameter. If the liquid is not
listed select Other and enter a propagation rate in metres/second.
This may be found in the back of the manual under Liquid Sound
Speeds.
QUICK START
yy-mm-dd hh:mm:ss
Select fluid type:
Water
Glycol/water 50/50
Lubricating oil
Diesel oil
Freon
Measure
Other (m/sec)