The measuring unit („mm“ and „inch“) and scale interval have to be
selected before setting the tolerance limits otherwise incorrect
measuring results may occur.
Configuration tolerance ranges: Example tolerance range 3 / Adjusting the tolerance limit –
Example: 10,00 ±0,1mm in the „Absolute – Mode“
1x long
Activating the SETUP – Menu in the
„Absolute – Mode“
2x short
„
tOL
“ will appear in the display
1x long
“1“ for tolerance range 1 will appear in the display
The last selected status will appear in the display /
Factory setting: 1
2x short
Switch to tolerance range “3“
1x long
„
tOL LO
“ will appear in the display for 2s
The last saved lower tolerance value will appear.
Adjusting
lower
tolerance limit (9,90mm)
Program: TOL