43
7
short
…..Reducing the first digit
8
1x short
Selecting the next digit
9
Repeat step 6-8 until the OFFSET value is totally entered.
10
1x long
For confirming and saving
11
Afterwards you will return to the SETUP -
Menu.
12
1x short
Quitting the SETUP – Menu
13
1x long
For confirming and saving.
The instrument returns to the last program
used
The instrument must be adjusted until the nominal value is reached!
If „
ERR 07
“ appears in the display the OFFSET value is out of the acceptable
range (
Technical Data). The instrument must be returned to the
manufacturer for checking.
You can select between „mm“ and „inch“.
The gage must be switched on. The SETUP – Menu can be activated in each program.
Zero point adjustment (OFFSET)
Changing the unit of measurement mm / inch (UNIT)