90
Performing User Calibration
4. By inputting correction coefficients
Operating Procedure
1
While the measurement screen is
displayed, press the [MENU] key.
The
setting screen
is displayed on the LCD screen.
2
Press the [UP] or [DOWN] key to select
[User Calib], and then press the [RIGHT]
or [ENTER] key.
MENU
Trigger mode 0-cal AUTO
Meas Cond
User Calib
Target
Option
System
Measure and Register
Copy and Register
Edit Correction Coef
Set ID
Delete Calib Data
3
Press the [UP] or [DOWN] key to select
[Edit Correction Coef], and then press
the [RIGHT] or [ENTER] key.
MENU
Trigger mode 0-cal AUTO
Meas Cond
User Calib
Target
Option
System
Measure and Register
Copy and Register
Edit Correction Coef
Set ID
Delete Calib Data
4
The screen for selecting a probe is
displayed. Press the [UP] or [DOWN] key
to select the probe to register, and then
press the [RIGHT] or [ENTER] key.
MENU
Trigger mode 0-cal AUTO
Meas Cond
User Calib
Target
Option
System
10
09
08
07
06
05
04
03
02
01
▶
Measure and Register
Copy and Register
Edit Correction Coef
Set ID
Delete Calib Data
5
The screen for selecting a calibration
channel is displayed. Press the [UP] or
[DOWN] key to select the channel to
register.
This screen allows the current data to be checked.
User Calib > Edit Correction Coef > Select CH
CH00
CH01
CH02
CH03
CH04
CH05
CH06
CH07
CH08
CH09
01
MENU
6
Press the [ENTER] key.
The screen for inputting correction coefficients is
displayed.
Содержание CA-410 Series
Страница 12: ......
Страница 40: ......
Страница 139: ......
Страница 140: ...En 9222 AA1J 12 2017 KONICA MINOLTA INC BIFBDA...