32
CA-P410
CA-P410H
Measurement area
Ø10 mm
Ø10 mm
Acceptance angle
±5°
±5°
Accuracy guaranteed measurement distance
30±5 mm
30±5 mm
Display range Luminance
0.0001 to 5000 cd/m
2
0.0001 to 30000 cd/m
2
Chromaticity
Displayed in 4 digits
Displayed in 4 digits
Luminance
Accuracy guaranteed luminance range
0.01 to 5000 cd/m
2
0.1 to 30000 cd/m
2
Accuracy (for white)
*1, *3
0.001 - cd/m
2
---
---
0.01 - cd/m
2
±2.5%
---
0.1 - cd/m
2
±2%
±2.5%
1 - cd/m
2
±2%
±2%
10 - cd/m
2
±1.5%
±2%
100 - cd/m
2
±1.5%
±1.5%
Repeatability
(2
σ
)
*1
AUTO
0.001 - cd/m
2
---
---
0.01 - cd/m
2
2%
---
0.1 - cd/m
2
0.60%
2%
1 - cd/m
2
0.20%
0.60%
10 - cd/m
2
0.10%
0.20%
100 - cd/m
2
0.10%
0.10%
Chromaticity Accuracy guaranteed luminance range
0.01 to 5000 cd/m
2
0.1 to 30000 cd/m
2
Accuracy (for white)
*1, *3
0.01 - cd/m
2
±0.006
---
0.1 - cd/m
2
±0.002
±0.006
1 - cd/m
2
±0.002
±0.002
10 - cd/m
2
±0.002
±0.002
100 - cd/m
2
±0.002
±0.002
In 100 cd/m
2
(for monochrome)
*2
100 cd/m
2
±0.003
±0.003
Repeatability
(2
σ
)
*1
AUTO
0.01 - cd/m
2
0.0070
---
0.1 - cd/m
2
0.0020
0.0070
1 - cd/m
2
0.0008
0.0020
10 - cd/m
2
0.0005
0.0008
100 - cd/m
2
0.0003
0.0005
Flick
er (CA
-310 Mode)
*6
Flicker
(Contrast)
Measurement luminance range
15 to 3000 cd/m
2
or more
90 to 18000 cd/m
2
or more
Range to be measured (flicker frequency)
0.25 to 65 Hz
0.25 to 65 Hz
Accuracy
30 Hz, AC/DC 10% sine wave ± 0.4%
± 0.4%
60 Hz, AC/DC 10% sine wave ± 0.7%
± 0.7%
Repeatability (2
σ
)
20 to 65 Hz,
AC/DC 10% sine wave
0.3%
0.3%
Flicker
(JEITA)
Measurement luminance range
15 to 3000 cd/m
2
or more
90 to 18000 cd/m
2
or more
Range to be measured (flicker frequency)
0.42 to 65 Hz
0.42 to 65 Hz
Accuracy
30 Hz, AC/DC 4% sine wave ±0.35 dB
±0.35 dB
30 Hz, AC/DC 1.2% sine wave ±0.35 dB
---
Repeatability (2
σ
)
30 Hz, AC/DC 4% sine wave 0.1 dB
0.1 dB
30 Hz, AC/DC 1.2% sine wave 0.3 dB
---
Specifications of Each Probe (2)
Содержание CA-410 Series
Страница 12: ......
Страница 40: ......
Страница 139: ......