PROCEDURE Setting Tilt Correction
Operating Procedure
Operation
Display
1) Turn on the machine, enter MEAS screen.
POWER
MEAS(1/3)
SHV
S
H
V
VA 273
°
18
′
15
″
HAR 150
°
18
′
15
″P1
DIST D/C 0SET HSET
2) Press ESC to enter status screen.
ESC
Home Page
Date 2015/07/28
Time 17:18:35
Ver. 20150415
Type KTS-460R6L
No. K76400
Job A:/JOB.JOB
MEAS MENU MEM CNFG
3) From the status screen press CNFG to enter the
configuration setting screen.
CNFG
Settings
(
1/2
)
1.Obs. Condition
2.Instr.Const
3.Date&Time
4.Comms Setup
5.Unit
6.Key Function
4) Select “1. Obs. Condition” and press ENT (or
press numeric key 1). Use ▲or▼ key to align the
cursor to the fourth line “Tilt crn”, use or to
set the tilt correction method. Then press ENT to
finish set. There are three options about tilt
correction: There are three options:
NO, X-ON and XY-ON.
“
1.Obs.
Condition
”
+
OK
+
▲
or
▼
+
or
Obs. Settings(1/2)
C&R crn: K=0.14
<
>
V.obs: Zenith
<
>
TiltCrn: OFF
<
>
PowerOff:OFF
<
>
Coord.: N-E-Z
<
>
5) Press ESC to return to setting screen.
ESC
Settings
(
1/2
)
1.Obs. Condition
2.Instr.Const
3.Date&Time
4.Comms Setup
5.Unit
6.Key Function