www.kolidainstrument.com
160
3) Press ▼ key to choose “2.V0/ Adjustment”, then
press ENT key(or press numeric key2), enter
Vertical index difference adjustment function.
“2.V0
Adjustment
”
+
ENT
Instr. Const
1. V0/Axis const.
2. V0 Adjustment
3. Collimation
4. Horizontal Axis
4) At positive position (Face left) sight at target,
press MEAS.
Face left,
sight at target
+
MEAS
V0 Adjustment
Step-1
V 90°21′49″
HR 185°47′42″
Vco 0°00′00″
SET RSET MEAS
5) Rotate telescope. At the reverse position (face
right) sight at the same target precisely, press
MEAS.
Face right
and sight at
target
+
MEAS
V0 Adjustment
Step-1
ZA 270°00′00″
HAR 5°50′15″
Vco 0°00′00″
SET RSET MEAS
6) After adjustment, it shows “Vco Complete!” on
the screen.
SET
V0 Adjustment
Vco Complete!
V 0°21′39″
HR 180°01′07″
Vco -0°10′55″
SET RSET MEAS
7) The display returns to Instrument constant menu
screen.
Instr. Const
1. V0/Axis const.
2. V0 Adjustment
3. Collimation
4. Horizontal Axis
Note:
1) Repeat the inspection steps to measure the Index Difference (
i
angle). If the Index difference
can not meet the requirement, you should check whether the three steps of the adjustment are right,
the sighting is right or not and etc. Then set again according to the requirement.
2) If Index Difference still can not meet the requirement after the repeated operation, the instrument
should be returned to service center for inspection and repair.
NOTE: The vertical angles shown in the Vertical Angle 0 Datum adjustment are only for
adjustment, don‟t use it as working result.