ADA5000 Automatic Distillation Analyzer 

Operation and Instruction Manual 

ADA5000-Manual 

14 -

 

Method 

Select method (ASTM D86, D1078, D850, DIN 
51751) 

 

Dist.Group 

Select distillation group (DG 0-3, DG 4, DG 
Solvents, DG 7,8, DG 6) 
 

Condenser 

Input condenser temperature. 
 

Receiver 

Input temperature to cool down the receiver 
compartment.  To warm the receiver chamber, 
select the box and enter the desired 
temperature. Note that cooling only works if the 
condenser setpoint temperature is within 0.5°C 
of the actual condenser temperature. 
 

Expected Vol. 

Input your expected volume. 
According to the standard, 5mL before this point 
heat capacity will not changed 
 

Preheat #1 

The first step of the pre-heating for the flask. 
This step influences mainly the time interval up 
to the first drop. 
 

Preheat Time 

Time in minutes for the adjusted first heating 
step (standard 3 minutes). 

 
Preheat #2 

Second step of the pre-heating capacity for the 
flask. This step influences mainly the rate of 
distillation of the first 2 mL. After 2mL the 
microprocessor of distillation analyzer will 
regulate the desired distillation rate. 
 

Opt. Heat #1 

If distillation run has finished, (min. 30 mL are 
necessary) microprocessor system of distillation 
analyzer will calculate optimized value. 
According to the distillation group a heating rate 
will be calculated to get a time for IBP of 300 sec 
in group 1-3, or 600 sec in group 4. 
 

Opt Time 

Optimized time for preheat step 1 will be 
calculated by microprocessor of distillation 
analyzer. 
 

Opt Heat #2 

If distillation run has finished, (min. 30 mL are 
necessary) microprocessor system of distillation 
analyzer will calculate optimized value to get 
desired distillation rate for the first 2 mL. 
 
Select „COPY“ to use optimized values for next 
distillation run. 
 

Extended 

 

 

Plate hole 

Select ceran plate (25mm, 32mm, 38mm, 
50mm) 

Flask Type 

Select flask type (125mL, 200mL) 

Distillation Rate 

Input distillation rate. According to the standard, 
a rate between 4 mL/min and 5 mL/min is 
requested. Use 4.5 mL/min. 

Final Heat Adjust 

If last 5 mL of distillation run are not in the time, 
correction of heating capacity upwards or 
downwards is possible. Heat capacity at 
constant heat (5mL before ‘Expected volume‘) is 
set to 100%. Correction of ±20% is possible. 
 

Stop At Vol. 

The distillation run can be interrupted at selected 
volume if the entire volume is not necessary. 
 

Stop At Temp. 

The distillation run can be interrupted at a 
selected temperature if the entire volume is not 
necessary. 

 

Содержание ADA5000

Страница 1: ...venue Bohemia New York 11716 1796 USA Toll Free 1 800 878 9070 US only Tel 1 631 589 3800 Fax 1 631 589 3815 http www koehlerinstrument com e mail info koehlerinstrument com Petroleum Testing Analysis...

Страница 2: ......

Страница 3: ...3 1 Packing List 3 3 2 Unpacking 3 3 3 Instrument Installation 3 3 4 Software Installation 3 4 Operation 7 4 1 Description of Instrument Controls 7 4 2 Instrument Operation 7 4 3 Communication Manage...

Страница 4: ...o are experienced in both petroleum products testing and instrument service to better understand your requirements and provide you with the best solutions You can depend on Koehler for a full range of...

Страница 5: ...e analyzer must be disconnected from the main power supply otherwise electrical hazards may be experienced Be extremely careful because of electrical shock and danger of personal injury Unit Design Th...

Страница 6: ...different models are K45603 K45703 TS 120VAC 5 50 60Hz 20A K45604 K45704 TS 230VAC 5 50 60Hz 10A Nitrogen Nitrogen must be available in order for the automatic fire extinguishing system to work Conne...

Страница 7: ...nt by clicking on the Yes button Confirm that the computer meets these minimum requirements and click on the Next button Enter the name and company information of the user To utilize the software as a...

Страница 8: ...ext button to continue Verify that all each box of the four components are checked and click on the Next button Select or create a folder the add program icons to Click Next to continue Review the pro...

Страница 9: ...arious paths please consult with Koehler technical service To change the units for temperature 1 and pressure 2 select the desired units by clicking on the corresponding radio buttons Registration 3 F...

Страница 10: ...amber 4 2 Instrument Operation 1 Power Turn on the main power switch to the unit Turn on the power switches on the PC and printer if applicable Note the internal computer on the K457XX TS models has a...

Страница 11: ...correct position Place the silicone stopper over the side arm of the distillation flask and then place it into the condenser tube 13 Move the heater carefully under the flask 14 Place the dropping pl...

Страница 12: ...e the status of an analyzer from the PC select the View option and then Unit Data Select unit 1 Select a connected analyzer to see all the important values of this analyzer Analog 2 Temperature shown...

Страница 13: ...sible to change the path of the database program form and results form For a normal installation you can use this standard directory If a PC is connected to a network it s possible to have the databas...

Страница 14: ...ults Form Select this item to change a result form Be careful when changing a form A backup of these forms can be found on the installation CD in directory Backup Results Comparison To compare up to 4...

Страница 15: ...ram Unit Select analyzer Load Select item to load a new program from the database Attention Program will be sent automatically to selected analyzer and overwrite an existing previous run If desired st...

Страница 16: ...Koehler Comment Sample Program Method ASTM D86 Distillation Group SG4 Condenser temp 40 0 C Receiver temp 30 0 C Expected volume 98 0 mL Preheat 1 29 1 Preheat time 3 0 min Preheat 2 28 2 Extended Fi...

Страница 17: ...finished min 30 mL are necessary microprocessor system of distillation analyzer will calculate optimized value According to the distillation group a heating rate will be calculated to get a time for...

Страница 18: ...fix temperatures are stored in a memo field of the database and are online available Program Extended Fix Temps Final Results All important values are shown online here That means if the IBP was detec...

Страница 19: ...since start is shown Cool Time Timer for cooling device Condenser Shows actual temperature of condenser tube Volume Shows actual volume of selected analyzer Receiver Shows actual temperature of recei...

Страница 20: ...wing sections 5 1 Flask Heater Compartment and Auto Fire Extinguisher The design of the flask heater compartment makes this instrument one of the safest automatic distillation units available The phys...

Страница 21: ...in personal injury or death If you have any questions about maintaining your equipment please do not hesitate to contact the Koehler technical service department 6 1 Routine Maintenance The Koehler A...

Страница 22: ...s been properly installed operated and maintained Koehler Instrument Company must be advised in writing of the malfunction and authorize the return of the product to the factory The sole responsibilit...

Страница 23: ...ADA5000 Automatic Distillation Analyzer Operation and Instruction Manual ADA5000 Manual 20 Notes...

Отзывы: