ADA5000 Automatic Distillation Analyzer 

Operation and Instruction Manual 

ADA5000-Manual 

9 -

 

 

 
To observe the data transfer process from the 
PC to the analyzer, select the “View” option and 
then “Data Transfer”.   
 

 

 
In this window the data transfer from PC to 
analyzer (blue color) and answer from analyzer 
(green color) sent back to PC can be observed. 
Unit 1 get the order “D“ and “64“ is the 
checksum. The answer from unit 1 follows and 
“BC“ is the checksum. The parameters which 
are set automatically by the configuration 
program are “9600-8N1“ and cannot be 
changed. 
 
If you select “File – File protocol“, then all the 
data transmission will be written into a file 
named “..\Koehler\bin\COMMREP.TXT“. In case 
of trouble with the communication, this file can 
be analyzed. Please don’t forget to close the file 
(“File – File protocol“). Please note: If 
“Commgr.exe” is seen, then the data 
communication to read a distillation run will be 
reduced. It is possible that distilled volume is 
different than seen on the result window of 
“distcntl.exe”. Use “commgr.exe” only for service 
and close it as soon as possible. 
 
To observe the status of an analyzer from the 
PC, select the “View” option and then “Unit 
Data”. 

 

 

 

Select unit 

(1)

Select a connected analyzer 

to see all the important values of this 
analyzer. 

 

 

Analog 

(2)

Temperature shown in °C of the 

flask, condenser tube, cooling device and 
receiver compartment can be seen. If the 
automatic dry point kit is installed and a 
thermocouple is connected, then the two 
temperatures, FeCo and compensation, can 
be observed. 

 

 

Digital/IO 

(3)

The LED’s show the status of 

all internal components. If an LED is “on“, 
then the component is running. If you left 
click with the mouse on one of these 
components, then you can switch it on or off 
for a short moment to check it. A login 
request screen will appear to change these 
parameters.  You will remain “logged in” until 
the software program is closed. 

 

 

Status 

(4)

In the upper section the internal 

information of the analyzer is shown and the 
lower section shows information during a 
distillation run. 

 

 

Adjust / Adjust Enable 

(5/6)

All calibration 

values are shown here. To change the 
calibration values, click on the “adjust 
enable” button 

(6)

. To perform an auto 

calibration on all temperature readings, click 
on “Run Adjustment” and follow the 
instructions in the pop-up boxes.  In order to 
perform the calibration, you will need a Pt–
100 simulator. 

 
 
 

Содержание ADA5000

Страница 1: ...venue Bohemia New York 11716 1796 USA Toll Free 1 800 878 9070 US only Tel 1 631 589 3800 Fax 1 631 589 3815 http www koehlerinstrument com e mail info koehlerinstrument com Petroleum Testing Analysis...

Страница 2: ......

Страница 3: ...3 1 Packing List 3 3 2 Unpacking 3 3 3 Instrument Installation 3 3 4 Software Installation 3 4 Operation 7 4 1 Description of Instrument Controls 7 4 2 Instrument Operation 7 4 3 Communication Manage...

Страница 4: ...o are experienced in both petroleum products testing and instrument service to better understand your requirements and provide you with the best solutions You can depend on Koehler for a full range of...

Страница 5: ...e analyzer must be disconnected from the main power supply otherwise electrical hazards may be experienced Be extremely careful because of electrical shock and danger of personal injury Unit Design Th...

Страница 6: ...different models are K45603 K45703 TS 120VAC 5 50 60Hz 20A K45604 K45704 TS 230VAC 5 50 60Hz 10A Nitrogen Nitrogen must be available in order for the automatic fire extinguishing system to work Conne...

Страница 7: ...nt by clicking on the Yes button Confirm that the computer meets these minimum requirements and click on the Next button Enter the name and company information of the user To utilize the software as a...

Страница 8: ...ext button to continue Verify that all each box of the four components are checked and click on the Next button Select or create a folder the add program icons to Click Next to continue Review the pro...

Страница 9: ...arious paths please consult with Koehler technical service To change the units for temperature 1 and pressure 2 select the desired units by clicking on the corresponding radio buttons Registration 3 F...

Страница 10: ...amber 4 2 Instrument Operation 1 Power Turn on the main power switch to the unit Turn on the power switches on the PC and printer if applicable Note the internal computer on the K457XX TS models has a...

Страница 11: ...correct position Place the silicone stopper over the side arm of the distillation flask and then place it into the condenser tube 13 Move the heater carefully under the flask 14 Place the dropping pl...

Страница 12: ...e the status of an analyzer from the PC select the View option and then Unit Data Select unit 1 Select a connected analyzer to see all the important values of this analyzer Analog 2 Temperature shown...

Страница 13: ...sible to change the path of the database program form and results form For a normal installation you can use this standard directory If a PC is connected to a network it s possible to have the databas...

Страница 14: ...ults Form Select this item to change a result form Be careful when changing a form A backup of these forms can be found on the installation CD in directory Backup Results Comparison To compare up to 4...

Страница 15: ...ram Unit Select analyzer Load Select item to load a new program from the database Attention Program will be sent automatically to selected analyzer and overwrite an existing previous run If desired st...

Страница 16: ...Koehler Comment Sample Program Method ASTM D86 Distillation Group SG4 Condenser temp 40 0 C Receiver temp 30 0 C Expected volume 98 0 mL Preheat 1 29 1 Preheat time 3 0 min Preheat 2 28 2 Extended Fi...

Страница 17: ...finished min 30 mL are necessary microprocessor system of distillation analyzer will calculate optimized value According to the distillation group a heating rate will be calculated to get a time for...

Страница 18: ...fix temperatures are stored in a memo field of the database and are online available Program Extended Fix Temps Final Results All important values are shown online here That means if the IBP was detec...

Страница 19: ...since start is shown Cool Time Timer for cooling device Condenser Shows actual temperature of condenser tube Volume Shows actual volume of selected analyzer Receiver Shows actual temperature of recei...

Страница 20: ...wing sections 5 1 Flask Heater Compartment and Auto Fire Extinguisher The design of the flask heater compartment makes this instrument one of the safest automatic distillation units available The phys...

Страница 21: ...in personal injury or death If you have any questions about maintaining your equipment please do not hesitate to contact the Koehler technical service department 6 1 Routine Maintenance The Koehler A...

Страница 22: ...s been properly installed operated and maintained Koehler Instrument Company must be advised in writing of the malfunction and authorize the return of the product to the factory The sole responsibilit...

Страница 23: ...ADA5000 Automatic Distillation Analyzer Operation and Instruction Manual ADA5000 Manual 20 Notes...

Отзывы: