Key Operator Instructions
5– 72
16FEB01 – 7E6174
6. Press DEV CAL to calibrate the Developer Pump, or FIX CAL to calibrate the
Fixer Pump.
Warning
• Gloves and protective eye wear must be worn when working with the processor
chemistry, especially when mixing the chemistry. See "Chemical Hazards" on
Page 1–9.
• If the processor chemicals get in the eyes, flush them thoroughly with large
amounts of water. If irritation persists, check with a doctor.
• Replenishment solutions are pumped quickly and may splash.
7. Carefully pull out the Processor until it locks into place.
8. Remove the Fixer Replenishment, or the Developer Replenishment hose and
place it into the Graduated Cylinder.
9. Hold the Graduated Cylinder over the Wash Section of the Processor so that spills
will not contaminate the processing solutions.
H186_8586AC
FIX
CAL
RETURN
CAL
DONE/
DEV
H186_1087AC
H186_1087ACA
Fixer
Replenishment
Hose
Developer
Replenishment
Hose
Graduated
Cylinder