Select SOLENOID ON ............................................................................ press ENTER
Select SOLENOID OFF ............................................................................ press ENTER
2
.
Exit the SERVICE PROGRAM
Press BACKSPACE 3 times
Select LEAVE COMPONENT TEST .................................................... press ENTER
Select QUIT ML2000 SERVICE MODE ............................................. press ENTER
Select Quit the program ........................................................................... press ENTER
REPLACEMENT OF THE SENSOR SUCKER BAR TILT B19
1
.
Switch off the ML2000 (ML2000 P).
2
.
Take off the TOP COVER, the REAR and the RIGHT-HAND PANEL.
3
.
Manually move the CASSETTE SUCKER BAR to the front.
4
.
Take out SENSOR B19. Mark the position of the WIRE TIES.
5
.
Install the new SENSOR.
6
.
Fix the harness with WIRE TIES. Use the marks from step 4 as reference.
FUNCTION TEST
1
.
Switch on the ML2000 (ML2000 P)
2
.
Test the function of SENSOR B19 with the SENSOR TEST.
Start the SERVICE PROGRAM
Select SERVICE MODE from the GLOBAL MENU ......................... press ENTER
ENTER SERVICE MODE MESSAGE is displayed .......................... press ENTER
UNIT DATA are displayed....................................................................... press ENTER
Select COMPONENT TEST .................................................................... press ENTER
Select SENSORS......................................................................................... press ENTER
Select WITH/WITHOUT SOUND ........................................................ press ENTER
SENSOR B19
figure 2-32
REPLACEMENTS
SM 3477
6/97
2-34
KODAK AG, Stuttgart