Specifications
B-4
September 2000
Power Outlets
A separate, dedicated power line with a 208 V, 15 amp dedicated receptacle is
required for the printer.
The outlets should meet the following standards:
Line Frequency
In most cases, the electric utility company is responsible for maintaining the
proper frequency of the alternating current in its distribution network. If the
building is supplied with power by its own motor-generator, make certain that the
governor that controls the frequency is capable of delivering power within the
required operating frequency range.
Item
Printer Standard
amperage
15 amp
wire size
14 gauge or larger
phase 1
(single)
receptacle
NEMA5—N6/15
impedance between ground and
neutral wires
less than 2 ohms
voltage drop across the circuit
breaker
less than 0.1 volts
distance between receptacle and
equipment
1.5 m (5.0 ft.)
or less
Содержание LED II PRINTER 20R - OPERATOR'S GUIDE
Страница 1: ...KODAK PROFESSIONAL LED II Printer 20R Operator s Guide Part No 6B6192 ...
Страница 6: ......
Страница 12: ......
Страница 16: ......
Страница 28: ......
Страница 70: ......
Страница 96: ......
Страница 116: ...Additional Calibration Information C 12 September 2000 Value Graph ...
Страница 117: ...Additional Calibration Information September 2000 C 13 Difference Graph ...
Страница 138: ......
Страница 160: ......
Страница 165: ......