4-12
A-61580 November 2007
Use the calibration target provided with your scanner.
1. Click
Calibrate
. The Calibrate dialog box will be displayed.
2. Select
Image
Chain
or
UDDS
from the drop-down box.
• If you select
Image Chain,
use the calibration target provided
with your scanner. Be sure to use a good, clean calibration target.
Place the calibration target in the input elevator face up and click
OK
.
IMPORTANT:If the White Background Accessory is installed, it
must be replaced with the black background strips,
and the scanner must be restarted prior to
calibration.
• If you select
UDDS,
place a 20 lb. or 75 g/m
2
bond sheet of paper
in the input elevator.
Do not use the calibration target
and click
OK
.
3. Click
Start
. While calibration is being performed a status will be
provided in the Status box.
4. Click
Done
when the calibration is successful.