20
Menu Display
Zero correction
Select module:
The module is in func-
tion check (HOLD) mode. The assigned
current outputs and relay contacts
behave as configured (BASE).
Press
enter
to confirm.
Zero correction:
Display of partial pressure, calibration
temperature and pressure.
Confirm with “OK”.
Select “Zero correction” calibration
method.
Press
enter
to confirm.
Zero Correction
For trace measurements below 500 ppb, the zero point should be calibrated.
If you want to perform a zero correction, then you should keep the sensor in
the calibration medium (eg, N2 or sulfite solution) until the measured value
has stabilized. This may well take several minutes. After that, you can start the
calibration process.
Calibration
Return
25.6°C
Module PH 3400-035
Module CONDI 3400-051
Info
80.3%Air
Module LDO 3400-170
Module LDO 3400-170
Return
25.6°C
80.3%Air
Automatic - Water
Automatic - Air
Product calibration Sat
Zero correction
Zero correction
Drift check running
Zero correction
25.6°C
80.3%Air
Partial pressure
Abort
0.30 mbar
OK
Calibration / Adjustment
Zero Correction
Calibration temperature
25.6 °C
Calibration pressure
1013 mbar