Parameter
Choices (default in bold print)
Input filter
Pulse suppression
Off
, Weak, Medium, Strong
Input filter
xxx s (
10 s
)
Sensor data
Measure in
Liquids
, Gases
Sensoface
On
/ Off
Sensor monitoring details
(all messages:
Off
, Maint. request, Failure)
Slope
Auto
Zero
Auto
Temperature
Auto
, Individual
Sensocheck
Off
, Auto
Sensor wear
Off,
Auto
Sensor operating time*
Off
, Individual
CIP counter*
Off
, Individual
SIP counter*
Off
, Individual
O
2
measurement with
CIP/SIP
Auto
, Individual
Auto-off function for O
2
measurement:
When the preset cut-off temperature is exceeded, the sensor
automatically stops the oxygen measurement to extend the life of
the membrane. The measured oxygen value is frozen, the temperature
continues to be measured.
Autoclaving counter*
Off
, Individual
Cal preset values
Product calibration
Sat
, Conc/p´(mbar)/p´(mmHg)
Cal concentration
mg/l, µg/l
, ppm, ppb
Cal timer
- Monitoring
Off
, Auto, Individual
- Cal timer
xxx h (
720 h
)
* Data is stored in the sensor