17
Menu Display
Automatic calibration in air
Drift check.
Display during calibration
• Partial pressure
• Calibration temperature
• Calibration pressure
If the signal or the measured tem-
perature fluctuate greatly, the
calibration procedure is aborted.
You must re-start the calibration.
If successful, place sensor in process,
exit calibration using softkey or
enter.
Calibration medium: Air
Select: First calibration
Enter relative humidity, e.g.:
• Ambient air: 50 %
• Bottled gas: 0 %
Enter cal pressure if “manual” has been
configured.
Start with softkey or
enter
Automatic - Air
Calibration medium: Air
25.6°C
80.3%Air
Start
Return
Relative humidity
0050 %
Input cal pressure
1013 mbar
Automatic - Water
Drift check running
25.6°C
80.3%Air
Partial pressure
Slope correction
214.0 mbar
Calibration tempe025.6°C
Calibration pressure
1013mbar
N
ot
e:
T
he displa
y ma
y v
ar
y depending on the devic
e v
ersion.