Product Calibration (Calibration with Sampling)
When the sensor cannot be removed – e.g. for sterility reasons – its slope can
be determined with “sampling”. To do so, the currently measured process value
is saved by the Protos. Directly afterwards, you determine a reference value
using a portable meter, for example. Then, you enter this reference value into
the process analysis system.
During calibration the module is in function check (HOLD) mode.
Current outputs and relay contacts of the module behave as configured
(BASE module).
NOTICE!
The reference value must be measured at temperature and pressure conditions
similar to those of the process.
Menu Display
Product calibration
Select module:
The module is in func-
tion check (HOLD) mode. The assigned
current outputs and relay contacts
behave as configured (BASE).
Press
enter
to confirm.
Select “Product calibration”
calibration method.
Sat (or Conc, p´) is preset in
Parameter setting / Cal preset values.
Press
enter
to confirm.
Calibration
25.6°C
Return
Info
72.5 %
Module LDO 3400-170
Module LDO 3400-170
25.6°C
Return
72.5 %
Calibration / Adjustment
Product Calibration (saturation/concentration/partial pressure [hPa, mmHg])
Preset in: Parameter setting / Cal preset values
Zero correction
Automatic - Water
Automatic - Air
Product calibration