Minimum Spans for Current Outputs
The CONDI 3400(X)-051 module is a measuring module. It does not provide
current outputs. Current outputs are provided by the BASE module (basic
device) or by communication modules (e.g. OUT module). The corresponding
parameters must be set there.
The minimum current span shall prevent that the resolution limit of the
measurement technology (± 1 count) is seen in the current.
CONDI 3400(X)-051 Module
S/cm
20 %, min. 100.0 µS/cm
% by wt
1.00
°C
10.0
g/kg
1.00
Ohm*cm
20 %, min. 100.0 ohms*cm
°F
10.0
Calculation Block COND/COND
Diff S/cm
20 %, min. 100.0 µS/cm
Diff °C
10.0
Diff Ohm*cm 20 %, min. 100.0 ohms*cm
RATIO
0.10
PASSAGE
10.0
REJECTION
10.0
DEVIATION
10.0