60
Select Curve, e.g. “Linear”:
The measured variable is represented
by a linear output current curve.
The desired range of the measured
variable is specified by the values for
“Start” and “End”.
See also: “Minimum span”
Current Outputs, Contacts, OK Inputs
Select menu: Parameter setting/Module BASE
Note: HOLD mode
Menu Display
Parameter setting
BASE module
• Select process variable
Output current
mS/cm
20
4
1999
0
Assignment of Measured Values: Start (4 mA) and End (20 mA)
mS/cm
20
4
700
500
[mA]
Example 1:
Range 0 ... 1999 mS/cm
Example 2: Range 500 ... 700 mS/cm
Advantage: Higher resolution in range of interest
[mA]
Configuring the Current Output
• Open parameter setting
• Enter passcode
• Select "Module BASE"
• Select “Output current ...”
Module BASE (Administrator)
19.2°C
Return
Contact K2 (NAMUR HOLD)
Contact K4 (NAMUR Failure)
Contact K3 (NAMUR maintenance)
0.020 mS/cm
Output current I2
Contact K1 (Limit)
Inputs OK1, OK2
Block
Output current I1 (Administrator)
OK
Variable
Output
Abort
Curve
Linear
°C
Trilinear
Function
Table
Output current I1 (Administrator)
OK
Curve
Output
Behavior during messages
Start
Abort
Variable
°C
g/kg
pH
°C
rH
Off
S/cm
End
Start
End
Output filter
500 700
19.2°C
0.020 mS/cm
19.2°C
0.020 mS/cm