76
Menu Display
Module diagnostics /
Sensor monitor / ServiceScope
Opening the diagnostics menu
From the measuring mode:
Press menu key to select menu.
Select diagnostics using arrow keys,
confirm by pressing enter.
Then select "Module CONDI".
The Diagnostics menu gives an over-
view of all diagnostics functions avail-
able. Messages set as “Favorite” can be
called up directly from the measuring
mode using a softkey.
To configure:
Parameter setting / System control /
Function control matrix.
Module Diagnostics
Internal function test (without Fig.).
Sensor Monitor
Shows the values currently measured
by the sensor. Important function for
diagnostics and validation!
(cf Maintenance)
Cal Record
Data of last calibration, suitable for
documentation to ISO 9000 and GLP
Menu selection
Return to meas
22.3 °C
Select:
[enter]
Lingua
Module CONDI 3400-051
22.3 °C
Return
Cal record
Sensor monitor
1.225 mS/cm
1.225 mS/cm
Set favorite
Module diagnostics
Sensor monitor
22.3 °C
1.225 mS/cm
Module Diagnostics
Module diagnostics, sensor monitor, cal record
Module diagnostics, Sensor
monitor, Cal record
Return
Resistance (c=1)
Conductance (c=1)
Temperature
100.0 Ohm
10.0 mS
25.6 °C
Cal record
Return
25.6°C
0.245 mS/cm
Active adjustment
Sensor model
Serial number
06/09/10 14:06
Cal mode
Cell factor
0077123
Product cal
6.2
SE670