72
Switching Parameter Sets via OK2
Parameter setting / System control / Function control matrix
Note: HOLD mode (Setting: BASE module)
Parameter Sets
2 complete parameter sets (A, B) can be stored in the analyzer.
You can switch between the parameter sets using the OK2 input.
The currently activated set can be signaled by a relay contact.
An icon in the measurement display shows which parameter set is active:
or
Menu Display
Parameter sets
Signaling Active Parameter Set via
Relay Contact
• Open parameter setting
• BASE module
• Select contact
• Usage: “Parameter set ...”.
Select Parameter Set (A, B) via OK2
Input
• Open parameter setting
• System control
• Function control matrix
• Select “OK2”
• Connect “Parameter set A/B”
Contact K3 (Administrator)
Abort
24.8°C
Contact type
OK
NAMUR maintenance!
Usage
ON delay
NAMUR HOLD
Limit value
Rinse contact
Parameter set B active
USP output
83.1 % Air
OFF delay
Note
The selection has no effect when working on SmartMedia card with
SW 3400-102.
24.8°C
Left softkey
83.1 % Air
Return
Connect
Function control matrix (Administrator)
Input OK2
Right softkey
Profibus DO 2
ParSet KI rec. Fav Unical
–
–
–
–
–
–