Maintenance
Maintenance
Sensor monitor, Temp probe adjustment
Note: HOLD mode active
Menu Display
Maintenance
Open Maintenance
From the measuring mode:
Press menu key to select menu.
Select maintenance using arrow keys,
confirm by pressing enter.
Passcode 2958 (The passcode can be
edited by the administrator.)
Then select Module CONDI.
Adjusting the temp probe
This function allows you to compen-
sated for the individual temperature
probe tolerance and the influence of
the lead resistances to increase accu-
racy of temperature measurement.
Adjustment may only be carried out
after the process temperature has been
precisely measured using a calibrated
reference thermometer!
The measurement error of the refer-
ence thermometer should be less than
0.1 °C. Adjustment without precise
measurement might result in consider-
able deviations of the measured value
display!
Sensor monitor
During maintenance, the sensor
monitor allows validation of the
sensor by immersing it in a known
solution, for example, and checking
the values measured.
Menu selection
Return to meas
25.6 °C
Select:
[enter]
Lingua
Return
25.6 °C
1.225 mS/cm
1.225 mS/cm
Adjust temp probe
Probe tolerance and lead adjustment
25.6 °C
0.225 mS/cm
Enter measured process temp
Installation adjustment
Abort
OK
Off
Process temperature
On
+
0
25.6°C
Sensor monitor
Resistance (c=1)
Conductance (c=1)
RTD
0.691 MΩ
1.225 mS
108.7 Ω
Temperature
25.6 °C
Sensor monitor
Adjusting the temp probe