6. Troubleshooting
Issue
Possible cause
Solution
Filter clogged
The pump was
switched off auto-
matically, because
Pmax was excee ded.
Exchange the filter
system
Pressure increase
Replace cartridge
Filter system inser ted
wrong
Exchange the filter
system
Unsteady baseline
Cartridge faulty
Replace cartridge
Filter clogged
Exchange the filter
system
7. Transport and storage
At the factory, the HPLC devices are carefully and safely packed for the
transport into special shipping boxes.
7.1 Packing the device
Original packaging: Ideally you should use the original transport packaging.
7.2 Transporting the device
Documents: If you want to return your device to KNAUER for repairs,
enclose the „
Service request form and decontamination report
“which can
be downloaded from our website.
7.3 Storing the device
Flushing solution: Pay attention that all hoses and capillaries have
been emptied or filled with flushing solution (e. g. isopropanol) before
storage. To prevent algae formation, do not use pure water.
Seals: Close all inputs and outputs with cap fittings.