SmartMix Instructions V5350
10
4. Installation
4.1 Mounting the SmartMix onto the pump
Phillips screwdriver, PH1 × 80
Legend:
1
Metal screw fitting as
blind plug
2
Assembly screws for
SmartMix mixer
3
Capillary to pump
4
Inlet for piston back-
flushing
5
Front panel
6
Pump head
7
Pressure sensor and
venting valve
8
Mixing chamber
9
Capillary to tube
Exterior view
Fig. 3:
Mounting SmartMix mixer on the pump
1
2
3
4
5
6
7
8
9
1.
Mount the SmartMix mixer to the front panel
5
of the pump by two
Phillips-head screws
2
.
2.
Tighten the screws with a Phillips screwdriver.
Device defect
Leakage can cause damage to the device .
Before startup, replace red or black PEEK blind plug in the LPG system
by metal screw fi tting (mounting kit F5351) .
Tool
Содержание SmartMix