SmartMix Instructions V5350
Basic safety instructions 5
2.4.2 Contamination by health-threatening solvents
Contamination with toxic, infectious or radioactive substances poses
a hazard for all persons involved during operation, repair, sale, and dis-
posal of a device.
All contaminated devices must be properly decontaminated by a spe-
cialist company or the operating company before they can be recom-
missioned, repaired, sold, or disposed .
2.4.3 Avoiding leakage
Risk of electrical shock or short circuit if solvents or other liquids leak into
the interior of the device. You can avoid a leakage through the following
measures:
Tightness: Visually check the device or system regularly for leaks.
Solvent tray: The use of a solvent tray prevents liquids get from the
bottles into the inside of the device.
Eluent lines: Install capillaries and hoses in such a way that, in case of a
leak, liquids cannot get into the interior of the devices underneath.
In case of leakage: Switch off the system. Only take the device into
operation if the cause of the leak has been resolved.
2.5 Specific environments
2.5.1 Earthquake-endangered areas
In earthquake-endangered areas, do not stack more than 3 devices on
top of each other. Otherwise there is risk of injury due to falling devices
or loose parts.
2.5.2 Explosive environment
Never use the system in potentially explosive atmospheres without
appropriate protective equipment. For more information, contact the
KNAUER Customer Support.
2.5.3 Cooling room
You may operate the device in a cooling room. To prevent condensation,
note the following instructions:
Allow the device to acclimatize for min. 3 hours before taking it into
operation.
After taking into operation, the device should stay switched on.
Avoid temperature fluctuations.
2.5.4 Wet room
The device must not be operated in wet rooms.