![Knauer Smartline Manager 5000 Скачать руководство пользователя страница 27](http://html.mh-extra.com/html/knauer/smartline-manager-5000/smartline-manager-5000_manual_1979833027.webp)
26
Interface Modul
can be independently programmed which enables the programming of
high pressure gradient system pumps.
Voltage range:
0 to +10 V
Minimum voltage step:
2.5 mV (12 bit)
(DAC resolution)
Maximum load resistance:
2 k
Ω
Digital Inputs (DIG.IN)
These four inputs are intended as a trigger (start signal) input for the start
signal that is distributed by the injection system. They work with contact
closure on mass and are compatible with TTL/CMOS or the Open
Collector. If the channels are used in separated systems, then every
channel can be separately started.
Maximum high level input voltage:
+15 V
Minimum high level input voltage:
+2.5 V
Maximum low level input voltage:
+1 V
Minimum low level input voltage:
-15 V
Maximum input current at V (in) = 0.5 V:
10 mA
Digital Outputs (DIG.OUT)
These four outputs are carried out in electromechanical relays with a
single pole double trough (SPDT). They can be selected as NO (normally
open; connection 2-3, see the back panel view) or NC (normally closed
connection 1-2, see the back panel view). Pulse or continuous signals
can be programmed. The length of a pulse contains approximately 1s;
the length of the continuous signal can be freely programmed.
Output type:
selectable NO or NC
Maximum switching voltage:
maximum 175 V DC
Maximum switching current:
maximum 0.25 A
RS-232 Interface
This interface is a 9-pin standard RS-232 interface (also known as a
serial interface or COM port) and is connected to a PC in order to realize
the software control of the interface.
Plug Charging
Pin 2
RxD
Pin 3
TxD
Pin 5
GND
Interface connection settings
Transfer rate:
19200 baud fix,
Data width:
8 bit
Stop bit:
1
Parity:
no
Connecting the Smartline Manager 5000 Interface Module
Power supply
The power supply is secured with the connection of the Smartline Pump
1000 (with a “PUMP” labeled socket on the back panel of the Smartline
Manager 5000). The required cable is included in the Smartline Manager
5000’s accessory kit.