Low Pressure Gradient (LPG) Block
9
100%. After this is completed, switch on the pump, so that the valve
can be opened.
Rinse the LPG block with the provided solvents by establishing a
connection on the pressure recorder’s vent capillary on the Smartline
Pump 1000 (see pos. {1.7} in Figure 1) using a syringe. Suction until there
are no more air pockets appear in the solution. Repeat this procedure for
all valves that will be used.
During operation the yellow “LEDs A to D” in front of the Smartline
Manager display the respective valve openings.
The accessory kit of the Smartline Manager 5000 includes a static
mixing chamber. It will be fixed at the front site of the pump, see
pos. {1.7} in Figure 1 on page 6. The mixing chamber enables an
optimum mixing of eluents in the range of 0.5 to 2.0 ml/min.
Technical Data
Eluent count:
up to 4
Maximum flow per channel:
50 ml/min
Gradient range:
0–100%
Gradient gradation:
1% steps
Control:
Through the Smartline Pump
1000 or ChromGate
®
or
EuroChrom
®
software