51
Maintenance and care
BlueShadow Pump 40P Instructions V7613
8.5.2.3 Installing the check valve
Component defect
Damage to components due to excessive tightening possible.
Observe the torque of the screw connection
Use 5 Nm torque for stainless steel fi ttings.
Use 1 Nm torque for PEEK fi ttings.
Insert the check valves in the direction of the flow. The notch of the check
valve points downward.
Insert the check valves in the direction of the flow. The arrow on the check
valve points upward.
The check valves have dried.
Procedure
Process
Figure
1.
Insert the check valves
1
so
that the notch of the check
valve
2
points downward.
2.
Manually screw in inlet and
out let fittings and tighten
them with a torque wrench
and the respective torque.
Fig. 21:
Check valve
Re-install the pump head.
Normal phase
Prerequisite
Next step
2
1