35
Operation
BlueShadow Pump 40P Instructions V7613
—
Flow rate value can be
changed. This creates
a second calibration
point which modifies the
scaling.
Time constant: Set the time
constant for the analog input
to be used for signal smoot-
hing (0.1-10.0 S). The greater
the value of the selected time
constant, the stronger the
applied signal smoothing.
Event check
Testing the terminal strip events.
Start-IN and Error-IN: Status
at the digital inputs.
Active only: Connects Event 1
to Event 8 and 24 V.
5.6 GLP menu
In the
GLP menu, statistical data on important parameters of the pump
are displayed.
Navigate through GLP menu
1.
Use the horizontal keys to navigate through the submenu.
2.
If no buttons are pressed the display returns to the main menu after a
short time.
3.
The menu contains no editable parameters.
5.7 Program menu
The menu
Program is used to list and create programs. The pump can
save up to 19 programs and up to 100 program lines with the following
functions:
Controlling the flow rate
Controlling the solvent composition
Sending control signals (Events)
Navigate through Program menu
1.
Open a program number of your choosing.
2.
Use the horizontal arrow keys to navigate through the two lines
of the display.
3.
To change a value press and change the value by the numeric pad.
4.
Move the cursor to the double arrow on the lower left line to navigate
to the next line or create a new line.
Note:
The maximum pressure for the program can be changed. It can
differ from the default one and is thereby only valid for this program.
Procedure
Procedure