32
Operation
BlueShadow Pump 40P Instructions V7613
5.5.3 Device setting parameters
Menu
Explanation
Figure
Cntl. Interface
Setting for external communi-
cation connection of the pump.
The connections are on the rear
side of the device.
RS232-11520, RS232-38400,
RS232-19200, RS232-9600
serial interface
LAN-DHCP: Automatic net-
work configuration
LAN manual: Manual net-
work-configuration (Port, IP
address, IP mask, Gateway)
Analog: Manual control of the
pump
Note:
After changing the
settings for external commu-
nication, the device should be
restarted.
Pr. limits
Settings for the minimum and
maximum pressure. The pump
stops automatically if the pres-
sure drops below the minimum
value or exceeds the maximum
value. The minimum pressure
value is used to ensure that the
pump does not run dry. If the
minimum is set to 0, the min-
imum system pressure is not
monitored.
Pump head
Pump head selection:
Automatic detection
10 ml
50 ml
Note:
Only possible if auto-
matic recognition of the pump
head does not function. The
selection is used as a basis for
calculating the flow rate and
determines the max. system
pressure. This value applies for
all operating modes and cannot
be exceeded.