45
Troubleshooting
BlueShadow Pump 40P Instructions V7613
System message
Not enough space
to store pro gram
Check the pump
Check the number of program lines – a maxi-
mum of 100 program lines are possible
Not supported
Change the entry.
P
Prg not compatible
with pump head!
Modify the program or replace the pump head.
Program does not
exist
Create and edit a program.
Program is running
Quit program or wait until program has been
com pleted.
T
The gradient
component is used!
First unload the program. The setup data can
only be edited when no program has been loa-
ded or started.
This link is used
in WakeUp
First quit or delete wakeup program (WakeUp),
then edit or delete link.
This program is
used in a link
First pause or delete link, then edit or delete
data by means of the chromatography software.
This program is
used in WakeUp
First quit or delete wakeup program (Wa-
keUp), then edit or delete data by means of the
chromatog raphy software.
Time already exists
Correct the time entry.
Time table line
is empty
Edit the program line.
Too many lines in
program
Check the number of program lines – a maxi-
mum of 100 program lines are possible.
U
Unknown pump
head type!
Check the pump head
Check whether a pump head with RFID
recogni tion has been installed
used!*/, CRC failed
Switch the device off and on. If the system mes-
sage appears again, notify the manufacturer for
technical support.
W
Wrong Line number Change the line index in the link.
Wrong link
Check Link no. – values from 1–10
are possible.
Correct the entry