22
Installation and initial startup
BlueShadow Pump 40P Instructions V7613
4.7.1 Firmware Wizard: Setting a static IP address
Note:
More information about LAN settings can be found in the Mobile
Control Software Instructions in the chapter "Firmware Wizard” (docu-
ment no. V6851-2).
1
Text box for serial
number of the
device
2
Setting IP address
manually
3
Text box for IP
address
4
Text box for subnet
mask & gateway
5
Confirm changes
Fig. 11:
Network settings in Firmware Wizard
The device is switched on.
Firmware Wizard is installed and running.
The connection between Firmware Wizard and the device has been
established.
Process
1.
In Firmware Wizard, click <Reset LAN Settings...>.
2.
The window <Device connection settings> opens. Enter serial
number of the device into the text field <Target device serial num-
ber>
1
.
3.
Select option <Use the following IP address>
2
.
4.
Enter the IP address into the text field <IP address>
3
.
5.
Optionally, adjust subnet mask and gateway
4
.
6.
Click <Reset Conn. Settings>
5
to accept changes.
7.
Restart the device (recommended).
The device is now accessible via the static IP address.
4.7.2 Firmware Wizard: Setting a dynamic IP address
The device is switched on.
Firmware Wizard is installed and running.
The connection between Firmware Wizard and the device has been
established.
1
2
3
4
5
Prerequisites
Procedure
Result
Prerequisites