Technical Data
Description
Data electrosurgical unit ME 102
Mains Connection
100 – 127 V / 220 – 240 V; 50 – 60 Hz
adjustable via solder bridge inside the unit
Power consumption
without RF output
approx. 16 VA
at max. output power
approx. 220 VA
Protection class
I
Degree of protection against
effect of water
IP X1
Classification acc. to German
Medical Devices Act
II b
Leakage currents LF / RF
acc. to IEC 601-2-2
Type
CF; defibrillation-proof
Nominal frequency
500 kHz
Modulation frequency
30 kHz
RF output variables
Current type: Macro mode
Output
Crest factor
Voltage
Cutting 1
max. 100 W ± 20 % at 600
Ω
1.8
1,200 V
ss
Cutting 2
max. 90 W ± 20 % at 600
Ω
2.3
1,600 V
ss
Contact coagulation
max. 70 W ± 20 % at 600
Ω
3.0
2,000 V
ss
Bipolar coagulation
max. 80 W ± 20 % at 100
Ω
1.9
440 V
ss
Current type: Micro mode
Output
Crest factor
Voltage
Cutting 1
max. 30 W ± 20 % at 600
Ω
1.8
600 V
ss
Cutting 2
max. 30 W ± 20 % at 600
Ω
2.3
800 V
ss
Contact coagulation
max. 25 W ± 20 % at 600
Ω
3.0
1,000 V
ss
Bipolar coagulation
max. 80 W ± 20 % at 100
Ω
1.9
240 V
ss
Operating mode
Intermittent INT 10 / 30 s corr. to 25 % duty factor
Mains fuses
110 – 127 V: T 2.0 AH; 220 – 240 V: T 1.0 AH; G 5 × 20 mm
Signal level
RF display: 55 dB(A)
Alarm: 65 dB(A)
Weight
Approx. 4.6 kg
Radio interference
suppression
Limit values acc. to EN 55011 and IEC 601-2-2
Radio interference immunity acc. to IEC 801
Dimensions W × H × D
256 × 97 × 320 mm
Environmental conditions for
transport and storage
Ambient temperature
Relative humidity
Air pressure
-25 °C to +70 °C (-13 °F to +158 °F)
10 – 100 %
500 – 1,060 hPa
Environmental conditions for
operation
Ambient temperature
Relative humidity
Air pressure
+10 °C to +40 °C
(-13 °F to +158 °F)
30 – 75 %
700 – 1,060 hPa
Compliant with 93/42/EEC