5.4
Working with the Active Electrode
Observe the following rules when using high-frequency surgery:
•
When simultaneously using electrosurgery and monitors on the same patient, use only
monitoring electrodes whose leads comprise protective resistors or RF reactors.
•
Needle electrodes must not be used for monitoring.
•
The active surgical electrode must not be used near ECG electrodes (minimum distance
15 cm).
•
The high-frequency power should be set to the lowest possible setting for the respective
application.
WARNING
Danger of burns from defective active electrodes!
Application of a current with high voltage, and in particular of a monopolar high voltage
coagulation current, may cause neuromuscular stimulation in the patient.
NOTICE
An insufficient effect while using the standard setting can, for example, be caused by poorly
applied neutral electrodes or insufficient contact in plug connections. Cables that are
damaged under the insulation or encrusted electrodes can also be a reason.
•
Replace defective parts if necessary.