Instructions for Use
Electrosurgical Unit ME 102
V. 3.1
15
4.6
Monopolar Operation
For the assignment of the numbering, see section 4.2 “Functions of the Control Elements,
Connections and Displays”, page 12.
4.6.1
Connection of the Neutral Electrode
Connect the neutral electrode to connection (5). You can connect either one-piece neutral
electrodes or two-surface electrodes that allow the monitoring of the contact to the patient.
If the neutral electrode is not connected, the red neutral electrode indicator LED (6) flashes. If
you try to activate the unit in a monopolar operating mode in this state via the finger or foot
switch, there will be an additional acoustic warning signal. The RF current cannot be activated.
This does not affect the bipolar coagulation current.
The indicator LED (6) goes dark after connecting a one-surface neutral electrode. Therefore
ensure that the neutral electrode is safely and fully attached to the patient. The indicator
LED (6) goes dark after reaching a safe application state when a multi-surface electrode (e. g.
KLS Martin TWIN-PAD) is correctly connected. As in this case individual run-in times are to be
expected, a lead time for the application must be considered.
WARNING
Risk of burns from improper handling of the neutral electrode!
A one-part neutral electrode cannot be monitored. In case of insufficient contact, there will be
no warning signal!