Instructions for Use
Electrosurgical Unit ME 102
30
V. 3.1
5.11
Additional Safety Notes
•
During operations on low-diameter body parts, bipolar technology can be used to avoid
unintentional coagulation in other areas.
•
Use of flammable anesthetics, nitrous oxide (N
2
O) and oxygen must be avoided. Sparking
on the active electrode will always occur when using electrosurgical devices. Combustible
substances that are used for cleaning or disinfection or as solvents must be evaporated
before application of electrosurgery.
•
There is the risk of accumulations of combustible liquid under the patient or in depressions
such as the navel or in body cavities such as the vagina. Liquid that has accumulated in
such areas must be removed before the electrosurgical unit may be used. Always be aware
of the possible presence of combustible endogenous gases. Materials saturated with
oxygen, such as cotton or gauze, may catch fire from the sparks that occur during the
intended use of the electrosurgical device. If possible, use non-flammable substances for
cleaning and disinfection.
•
The unit may be combined with other units only by the manufacturer or with the
manufacturer's consent.
•
The electrosurgical unit may interfere with other electromedical devices.