025-0169-311 Revision: B
Page 24 of 52
PARAMETER
Measurement Window
Available Modes
Available Settings
STATE
Status Description
Status
Description
Status
Description
Range
40–95ºC
Increment
1 ºC
SET
TEMP
Default*
90 ºC
Range
30 s–20 min
Increment
00:30 s
IDL MODE
ON
Temperature of IDL and Thermocouple Probe are
displayed. Impedance measurement for IDL is
displayed. Time and Power are not displayed
(dashed lines replace values). If the Thermocouple
Probe is disabled via IDL SETTINGS state,
accessible from the ADVANCED SETTINGS mode,
that measurement is greyed out and the value
replaced with dashed lines.
SAVE
SETTINGS
While in the READY
state, other operating
modes are not
available.
The present
SETTINGS can be
saved as defaults by
holding down the
SAVE SETTINGS key.
TIME
Default*
16:30 min
* Defaults are factory-set, and are overwritten when the SAVE SETTINGS function is used
†
Time Range set by user does not include factory-set PRE-TREATMENT COOLING Time (45 s) or POST-TREATMENT COOLING Time (10
seconds)