Figure 7-17 RFA Mode – TREATMENT ON Display
Figure 7-18 IDL Mode – ON Display (Secondary Thermocouple Enabled)
Displays and Indicators:
PARAMETER
STATUS
DESCRIPTION
Stimulus Output ON
Stimulus pulses are delivered during the ON state.
RF/Stimulus Output
RF Output ON
RF power is delivered during the ON state. SET TEMP/TEMP
LIMIT/WARNING LIMIT and treatment TIME remain modifiable
during RF power delivery.
COOLED RF AND TRANSDISCAL - RF power is delivered during
the TREATMENT ON state. If the procedure is stopped while in
TREATMENT ON state, the Generator enters POST-TREATMENT
COOLING state if enabled in ADVANCED SETTINGS mode.
FAULT Indicator
OFF
The FAULT indicator is used only in the POST and SYSTEM FAULT
states.
RF/Stimulus Output
ON/OFF Indicator
ON
The indicator is illuminated during Stimulus pulse/RF power delivery.
ON
RF audio output is generated for RF power delivery.
RF Audio Output
None
RF audio output is not generated during Stimulus pulse delivery.
Impedance Related
Audio Output
OFF
Impedance related audio output is not generated during the ON state.