025-0169-311 Revision: B
Page 7 of 52
•
The mains power cord of the Generator must be connected to a properly grounded receptacle. Extension cords and/or
adapter plugs must not be used.
•
Do not wrap instrument cable around metal objects. Wrapping cables around metal objects may induce hazardous
currents.
•
The cables to the surgical electrodes should be positioned in such a way that contact with the PATIENT or other leads
is avoided. Temporarily unused ACTIVE ELECTRODES should be stored in a location that is isolated from the
PATIENT.
•
Avoid high frequency output settings where the Maximum Output Voltage may exceed the rated accessory voltage
•
The output power should be as low as possible for intended purpose (Applies to Manual modes)
•
The application of the Neutral Electrode and its connections to be checked before selecting a higher output power
•
For surgical procedures where the high frequency current could flow through parts of the body having a relatively
small cross sectional area, the use of BIPOLAR techniques may be desirable in order to avoid unwanted tissue
damage.
•
Perform regular inspections of all accessories, including electrosurgical cables and probes, for damage to insulations
•
Associated Equipment and Active Accessories should be selected that have a Rated Accessory Voltage equal to or
greater than the Maximum Output Voltage
•
For information on the connection and disconnection of detachable parts and accessories, refer to the Instructions for
Use for the corresponding probe or cable
•
Care should be taken to avoid the danger of ignition of endogenous gases
•
The PMG needs special precautions regarding EMC and needs to be installed and put into service according to the
EMC information provided in this document.
•
Electronic equipment, including portable and mobile RF communications equipment, can affect the operation of the
PMG. Operating non-essential equipment in the vicinity of the PMG should be avoided, if possible.
•
The PMG should not be used adjacent to or stacked with other equipment. If the PMG must be operated adjacent to
or stacked with other equipment, the PMG should be observed to verify normal operation in that configuration.
•
Use of accessories, transducers and cables other than those specifically approved by Baylis for use with the PMG
may result in increased electromagnetic emissions or decreased electromagnetic immunity of the PMG.
•
In Multi-RF Auto Temp mode, maintain a minimum distance of 10 mm between probe tips and ensure that the probe
tips do not touch.
•
When using the PMG with the pump unit accessory (TDA-PPU) refer to the pump IFU for information on proper
setup and operation
4
Installation
Inspect the Generator for any signs of physical damage to the front panel, chassis or cover. If any physical damage is
found, DO NOT USE THE GENERATOR. CONTACT Baylis Medical for a replacement. Baylis Medical must
approve all returns.
4.1
Preparing The Generator For Use
•
The Generator may be placed on a mounting cart or on any sturdy table or platform.
•
Provide at least four to six inches (10-15 cm) of space behind the rear panel of the Generator for forced air-cooling.
Do NOT obstruct the vents on the underneath of the Generator. Under continuous use for extended periods of time, it
is normal for the top and rear panel to be warm.
4.2
Mains Power Cord
•
The Generator is shipped with an approved hospital-grade mains power cord.
•
Do not use extension cords or three-prong to two-prong adapters. The mains power cord assembly should be
periodically checked for damaged insulation or connectors.
4.3
Generator Cleaning And Disinfection Instructions
•
Use a mild detergent and damp cloth to clean the Generator cover, front panel, and power cable. The Generator
cannot be sterilized. Do not allow fluids to enter the chassis. The Generator may be disinfected using a standard
hospital alcohol solution applied with a cloth.
•
Do not spray or pour liquids directly on the Generator.