Keysight E1463A 32-Channel Form C Switch Service Guide
29
Performance Verification
Verification Tests
(Continued on Next Page)
290OUTPUT @Sw;"*RST"
300OUTPUT @Sw;"*OPC?"
310ENTER @Sw;Opc
320IF Opc<>1 THEN GOTO 300
330IF J=1 THEN
340OUTPUT @Sw;"CLOS (@"&Cc$&"00:"&Cc$&"31)"
350END IF
360PRINT TABXY(1,1),"Channels 00-31 "&Path$(J)& " measurements"
370PRINT TABXY(1,3),"1. Connect DMM Sense and Input HI leads to test
fixture "&Path$(J)
380PRINT TABXY(1,4),"2. Connect DMM Sense and Input LO leads to test
fixture C"
390DISP "Press Continue when connections are complete"
400PAUSE
410CLEAR SCREEN
420OUTPUT @Dmm;"TRIG SGL"! Check for stuck relays
430ENTER @Dmm;Meas
440IF Meas <1.E+6 THEN
450PRINT "Stuck relay found"
460PRINT "Repair problem and re-run test"
470STOP
480END IF
490FOR I=0 TO 31
500IF J<1 THEN
510IF I <10 THEN
520OUTPUT @Sw;"CLOS (@"&Cc$&Ph$&VAL$(I)&")"
530ELSE
540OUTPUT @Sw;"CLOS (@"&Cc$&VAL$(I)&")"
550END IF
560ELSE
570IF I <10 THEN
580OUTPUT @Sw;"OPEN (@"&Cc$&Ph$&VAL$(I)&")"
Содержание VXI Bus 75000 C Series
Страница 1: ...Keysight E1463A 32 Channel 5 Amp Form C Switch Service Manual 75000 Series C...
Страница 2: ......
Страница 6: ...6...
Страница 18: ...20 Keysight E1463A 32 Channel Form C Switch Service Guide General Information Recommended Test Equipment...
Страница 42: ...44 Keysight E1463A 32 Channel Form C Switch Service Guide Verification Tests Performance Test Record...
Страница 52: ...54 Keysight E1463A 32 Channel Form C Switch Service Guide Replaceable Parts Replaceable Parts List...
Страница 61: ......