14
Keysight E1463A 32-Channel Form C Switch Service Guide
General Information
Safety Considerations
GROUNDING REQUIREMENTS. Interruption of the
protective (grounding) conductor (inside or outside the
mainframe) or disconnecting the protective earth terminal
will cause a potential shock hazard that could result in
personal injury. (Grounding one conductor of a
two-conductor outlet is not sufficient protection.)
IMPAIRED PROTECTION. Whenever it is likely that
instrument protection has been impaired, the mainframe
must be made inoperative and be secured against any
unintended operation.
REMOVE POWER IF POSSIBLE. Some procedures in this
manual may be performed with power supplied to the
mainframe while protective covers are removed. Energy
available at many points may, if contacted, result in
personal injury. (If maintenance can be performed without
power applied, the power should be removed.)
USING AUTOTRANSFORMERS. If the mainframe is to be
energized via an autotransformer (for voltage reduction)
make sure the common terminal is connected to neutral
(that is, the grounded side of the main’s supply).
CAPACITOR VOLTAGES. Capacitors inside the mainframe
may remain charged even when the mainframe has been
disconnected from its source of supply.
USE PROPER FUSES. For continued protection against fire
hazard, replace the line fuses only with fuses of the same
current rating and type (such as normal blow, time delay,
etc.). Do not use repaired fuses or short-circuited
fuseholders.
SHOCK HAZARD. Only service-trained personnel who are
aware of the hazards involved should install, remove, or
configure the Keysight E1463A 32-Channel Form C Switch
Module . Before you remove any installed module,
disconnect AC power from the mainframe and from other
modules that may be connected to the Switch Module.
CHANNEL WIRING INSULATION. All channels that have a
common connection must be insulated so that the user is
protected from electrical shock. This means wiring for all
channels must be insulated as though each channel carries
the voltage of the highest voltage channel.
Содержание VXI Bus 75000 C Series
Страница 1: ...Keysight E1463A 32 Channel 5 Amp Form C Switch Service Manual 75000 Series C...
Страница 2: ......
Страница 6: ...6...
Страница 18: ...20 Keysight E1463A 32 Channel Form C Switch Service Guide General Information Recommended Test Equipment...
Страница 42: ...44 Keysight E1463A 32 Channel Form C Switch Service Guide Verification Tests Performance Test Record...
Страница 52: ...54 Keysight E1463A 32 Channel Form C Switch Service Guide Replaceable Parts Replaceable Parts List...
Страница 61: ......