Keysight E1463A 32-Channel Form C Switch Service Guide
27
Performance Verification
Verification Tests
–
Send CLOS (@nn00) to close channel 0, where nn = card # (typically 01)
3
Repeat for Channels 1 through 31
–
Repeat step 2 for channel 1 through 31 C to NC
–
Use OPEN (@nncc) and CLOS (@nncc), where nn = card # and cc =
channel # (omit leading zeroes in nn)
Example: Closed Channel Resistance Test
This example performs a Closed Channel Resistance Test to measure Channels 0
through 31 relay contact resistances. If the relay contact resistance for a channel
is >2.0
?
the program prints a message indicating which channel has failed the
test. Use this list in Chapter 4 when troubleshooting a failing relay.
Figure 1-6
Common to NC Test Connections