![Keysight VXI Bus 75000 C Series Скачать руководство пользователя страница 26](http://html1.mh-extra.com/html/keysight/vxi-bus-75000-c-series/vxi-bus-75000-c-series_service-manual_1961145026.webp)
28
Keysight E1463A 32-Channel Form C Switch Service Guide
Verification Tests
Performance Verification
10!RE-SAVE "CLOS_TEST"
20 ASSIGN @Dmm TO 722
30 ASSIGN @Sw TO 70915
40 DISP CHR$(129)
50 DIM Result(1,31),Path$(1)[2]
60 DATA NO,NC
70 READ Path$(*)
80 Cc$="01"! Card number
90 Ph$="0"! Place holder
100!
110!Start test
120!
130CLEAR SCREEN
140PRINT "Install Component Assembly and Test Fixture "
150PRINT
160PRINT " 1. Turn Mainframe and Keysight 3458A DMM power OFF"
170PRINT " 2. Connect GPIB Cable between mainframe and DMM"
180PRINT " 3. Install E1463A Component Assembly into Mainframe"
190PRINT " 4. Attach Test Fixture to Component Assembly"
200PRINT " 5. Turn Mainframe and DMM power ON "
210PRINT " 5. Press Continue when ready to begin testing "
220PAUSE
230CLEAR SCREEN
240!
250! Start the measurements
260!
270FOR J = 0 TO 1
280OUTPUT @Dmm;"PRESET NORM;FUNC OHMF"
This test assumes that the module is configured without
protection varistors, fuses, or resistors and that jumpers JM1
through JM8 are installed on the Keysight E1463A Component
Assembly (factory setting). Refer to the Keysight E1463A User’s
Manual for details and locations of user installed protection
circuits
Содержание VXI Bus 75000 C Series
Страница 1: ...Keysight E1463A 32 Channel 5 Amp Form C Switch Service Manual 75000 Series C...
Страница 2: ......
Страница 6: ...6...
Страница 18: ...20 Keysight E1463A 32 Channel Form C Switch Service Guide General Information Recommended Test Equipment...
Страница 42: ...44 Keysight E1463A 32 Channel Form C Switch Service Guide Verification Tests Performance Test Record...
Страница 52: ...54 Keysight E1463A 32 Channel Form C Switch Service Guide Replaceable Parts Replaceable Parts List...
Страница 61: ......