5
Characteristics and Specifications
Digital Multimeter Specifications
162
U3606B User’s Guide
[5] Specifications stated are for 4-wire low-resistance measurements. The test current is sent from the FORCE terminals and
the resistance is measured by the SENSE terminals.
The contact strength may influence the measuring result significantly. Ensure that the connection of the test point is firm to
avoid resistance due to contact leads.
The accuracy is specified after source compensation due to environment temperature changes. Initiate the compensation
by exiting and entering the Lo-
Ω
function or by disabling and enabling the output.
The measuring current will be reduced automatically when the product of the test current and resistance exceed 7.5 V. Refer
to the test current and resistance as shown below:
[6] Specifications stated are for the voltage measured at the input terminals only. The test current (1 mA) is typical. Variation in
the current source will create some variation in the voltage dropped across a diode junction.
[7] Specifications stated are for open test lead measurements and film capacitor or better using the Null math operation.
Test current
Maximum test
resistance
Test current
Maximum test
resistance
4 mA
<1200
Ω
8 mA
<938
Ω
5 mA
<1200
Ω
9 mA
<834
Ω
6 mA
<1200
Ω
10 mA
<750
Ω
7 mA
<1072
Ω
-
-
N O T E
For the total measurement accuracy, add the probe error. The contact strength will
significantly influence to the measuring result. Ensure proper contact at the test point you
want to measure.
Содержание U3606B
Страница 1: ...Keysight U3606B Multimeter DC Power Supply User s Guide ...
Страница 2: ...THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 3: ...U3606B User s Guide I ...
Страница 16: ...XIV U3606B User s Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 25: ...Introduction 1 Preparing the U3606B U3606B User s Guide 7 Adjust the carry handle 1 2a 3 2b ...
Страница 27: ...Introduction 1 Preparing the U3606B U3606B User s Guide 9 Figure 1 3 Rack mount dimensions ...
Страница 46: ...1 Introduction The U3606B in Brief 28 U3606B User s Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 120: ...3 DC Power Supply Operation Square Wave Operation 102 U3606B User s Guide Enable the output 7 ...
Страница 125: ...DC Power Supply Operation 3 Sweep Functions U3606B User s Guide 107 Connect the load 1 LOAD 3 2 ...
Страница 172: ...4 System Related Operation Remote Operation 154 U3606B User s Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...