Characteristics and Specifications
5
Digital Multimeter Specifications
U3606B User’s Guide
161
Capacitance
[7]
1.000 nF
0.001 nF
0.75 µA
current
source
-
-
2.0 + 0.8
0.02 + 0.001
10.00 nF
0.01 nF
0.75 µA
-
-
1.0 + 0.5
0.02 + 0.001
100.00 nF
0.1 nF
8.3µA
-
-
1.0 + 0.5
0.02 + 0.001
1.000 µF
0.001
μ
F
83 µA
-
-
1.0 + 0.5
0.02 + 0.001
10.00 µF
0.01
μ
F
83 µA
-
-
1.0 + 0.5
0.02 + 0.001
100.0 µF
0.1
μ
F
83 µA
-
-
1.0 + 0.5
0.02 + 0.001
1000 µF
1
μ
F
0.83 mA
-
-
1.0 + 0.5
0.02 + 0.001
10000 µF
1
μ
F
0.83 mA
-
-
2.0 + 0.5
0.02 + 0.001
[1] 20% over range on all ranges, except for 20 mV
dc
, 1000 V
dc
, and 3 A
dc
range.
[2] Relative to calibration standards.
[3] Any current measurement greater than 500 mA will have a temporary thermo-effect. If you wish to measure a lower current
or offset current immediately after a high current measurement, ensure that the U3606B has cooled down.
[4] Specifications stated are for 2-wire resistance measurements using the Null math operation. Without Null, add a 0.2
Ω
error.
To eliminate the noise interference, which might be induced due to the test leads, a shielded test cable is recommended for
measuring resistance above 100 k
Ω
.
Table 5-1
DC accuracy specifications ± (% of r % of range) (continued)
Function
Range
[1]
Resolution
Test current
or burden
voltage
24 Hours
[2]
23 ºC ± 1 ºC
90 Days
23 ºC ± 5 ºC
1 Year
23 ºC ± 5 ºC
Temperatur
e
coefficient
0 ºC to 18 ºC
28 ºC to 55
ºC
Содержание U3606B
Страница 1: ...Keysight U3606B Multimeter DC Power Supply User s Guide ...
Страница 2: ...THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 3: ...U3606B User s Guide I ...
Страница 16: ...XIV U3606B User s Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 25: ...Introduction 1 Preparing the U3606B U3606B User s Guide 7 Adjust the carry handle 1 2a 3 2b ...
Страница 27: ...Introduction 1 Preparing the U3606B U3606B User s Guide 9 Figure 1 3 Rack mount dimensions ...
Страница 46: ...1 Introduction The U3606B in Brief 28 U3606B User s Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 120: ...3 DC Power Supply Operation Square Wave Operation 102 U3606B User s Guide Enable the output 7 ...
Страница 125: ...DC Power Supply Operation 3 Sweep Functions U3606B User s Guide 107 Connect the load 1 LOAD 3 2 ...
Страница 172: ...4 System Related Operation Remote Operation 154 U3606B User s Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...