Operation and Features
Selecting a Range
60
U3606B User’s Guide
N O T E
•
Autoranging is selected at default factory power-on and after a remote
reset.
•
Manual ranging — If the input signal is greater than can be measured
on the selected range, the multimeter provides these overload
indications: “+/– OL” from the front panel or “
±9.9E+37
” from the
remote interface.
•
For frequency, pulse width, and duty cycle measurements, ranging
applies to the input voltage or input current of the signal, not its
frequency.
•
The range is fixed for diode (1 V
dc
range) tests.
•
The U3606B remembers the selected ranging method (auto or manual)
and the selected manual range for each measurement function.
•
Autorange thresholds — The U3606B shifts ranges as follows:
Down range at < 10% of current range
Up range at > 120% of current range
Содержание U3606B
Страница 1: ...Keysight U3606B Multimeter DC Power Supply User s Guide ...
Страница 2: ...THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 3: ...U3606B User s Guide I ...
Страница 16: ...XIV U3606B User s Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 25: ...Introduction 1 Preparing the U3606B U3606B User s Guide 7 Adjust the carry handle 1 2a 3 2b ...
Страница 27: ...Introduction 1 Preparing the U3606B U3606B User s Guide 9 Figure 1 3 Rack mount dimensions ...
Страница 46: ...1 Introduction The U3606B in Brief 28 U3606B User s Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 120: ...3 DC Power Supply Operation Square Wave Operation 102 U3606B User s Guide Enable the output 7 ...
Страница 125: ...DC Power Supply Operation 3 Sweep Functions U3606B User s Guide 107 Connect the load 1 LOAD 3 2 ...
Страница 172: ...4 System Related Operation Remote Operation 154 U3606B User s Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...